49. Balance RTD bridge

Circuit

Loading simulator…

Design Approach

An RTD’s resistance changes by only a few tens of ohms over the test range. A balanced two-branch bridge converts this small resistance change into a measurable differential voltage while keeping RTD current below 1 mA.

How the Circuit Works

  1. Balance at the reference temperature. The supplied RTD1 is 100 Ω at 0°C, matching fixed R_REF = 100 Ω. The solution adds equal upper resistors, R_L = 3.3 kΩ and R_R = 3.3 kΩ, creating two identical divider ratios at 0°C.
  2. Understand the zero output. At 0°C, both bridge midpoints are approximately 3.3×1003300+100≈97mV\frac{3.3 \times 100}{3300 + 100} \approx 97\,\mathrm{mV}. The differential meter V_DIFF reads approximately 0 V because both branch voltages match.
  3. Respond to heating. As temperature rises, RTD1 resistance increases while R_REF stays fixed. The RTD-side midpoint rises relative to the reference side. Around 100°C, a nominal 100 Ω RTD reaches approximately 138.5 Ω, giving a differential of about 36 mV, above the 30 mV requirement.
  4. Check sensor current. At 0°C, the RTD branch current is 3.33300+100≈0.971mA\frac{3.3}{3300 + 100} \approx 0.971\,\mathrm{mA}, below 1 mA. Increasing RTD resistance reduces it further.

Verify the Circuit

Check RTD1 at 0°C, 50°C, and 100°C. Expect V_DIFF near 0 mV at 0°C, about 18 mV at 50°C, and at least 30 mV at 100°C, with RTD current ≤1 mA throughout.

Key Learnings

  • A bridge compares a sensor against a matched reference, exposing small resistance changes as differential voltage.
  • Large upper resistors limit sensor current but also reduce the available differential signal.